9. Connect gauge through comm. box to the computer, wait for 4 (strain gauge), and 5
(quartz gauge) LED flashes on comm. box and download data to the hard drive (See
software operational manual).
CAUTION: DO NOT DISCONNECT GAUGE DURING DATA TRANSMISSION.
10. Remove old o’rings from I/O connector side of the gauge.
11. Wipe clean the threads and o’ring surfaces and inspect visually all threads for signs of
damage, galling and etc.
12. Install two new o’rings under the battery housing, (C-ring for 1-1/5” K10 SS gauge).
13. Lube threads with Kuster high temp lube.
14. Program tool for 2 point DWT.
15. Install battery and battery housing.
16. Run 2 point DWT.
17. Download the data from the gauge.
18. Review, record & file 2 point DWT.
19. Fill transducer port with fresh oil and install buffer tube assembly onto it.
20. Install tandem sub and bull nose.
DISASSEMBLE/ ASSEMBLE K10 STRAIN 1-1/4” AND K10 QUARTZ 1-1/4” GAUGES IN
ORDER TO CHANGE O’RINGS UNDER THE ELECTRONIC HOUSING
The frequency of changing o’rings under an electronic housing should be dependent on well fluid
types, temperature and gauge usage intensity. With high content H2S we recommend changing
these o’rings every 6 months or 1000 hours in well, which ever comes first.
The frequency of changing o’rings under battery housing is up to the tool operator discretion.
They should be replaced at the first evidences of wear.
All Kuster electronic gauges are furnished with Parker Company Viton o’rings, which cover the
majority of well fluids in the field, however the last decision on which type of o’rings to use is up to
the operator of the gauge.
A wrong choice could lead to tool malfunctioning and denial of warranty. Consult with the o’ring
manufacturer, if you are not sure that these o’rings are suitable for your applications. It is mainly
a concern for the well fluids with high content of H2S.
Electronic housing disassembly:
1. Unscrew slotted nut from the top of the LEMO connector with a fork wrench, which
comes with accessories.
2. Push connector just below the keyway cutout in the housing.
3. With the connector being below the keyway, carefully unscrew electronic housing from
the pressure sub.
4. Remove old o’rings.
5. Check thread and o’ring grooves and lubricate them with Kuster lube grease.