TM 11-6625-2611-12
Table 3-1. Operator’s Controls, Indicators, and Connectors (Continued)
Control, indicator,
or
connector
PRF SELECT (3-position) RANGE switch
PRF SELECT MULT control
MEASUREMENT PRF (3-position) RANGE switch
MEASUREMENT meter
CHAL SUB PULSE (4-position rotary) SELECT
switch
CHAL SUB PULSE (6-position rotary) POSITION
SELECT switch
CHAL SUB PULSE POSITION VARY control
CHAL MODE SELECT (9-position rotary) switch
Function
IN/OUT jacks. Peak power of signal applied to
LOW PWR MEAS IN jack can also be deter-
mined in conjunction with conversion chart
packed in the front panel cover and shown in
figure 3-2. This chart is used to convert meter in-
dication to actual power being measured.
FREQ
Test set is used to measure frequency of input rf
signal. Position of video signal with respect to
frequency markers on MEASUREMENT MKRS
OUT jack can be used to determine input signal
frequency. MEASUREMENT meter indicates
PRF of reply video from transponder under test.
Selects range of prf of challenge video generated by test set. Actual prf
within range is set by PRF SELECT MULT control. When CHAL MODE
SELECT switch is set to any position, prf is as indicated below. When
PRF SELECT switch is set to X1/2, prf is halved.
Switch position
Function
X10
Sets prf to range of 10 to 110 Hz.
X100
Sets prf to range of 100 to 1100 Hz.
X1K
Sets prf to range of 1000 to 11,000 Hz.
Adjusts challenge video prf within range selected by PRF SELECT
RANGE switch.
Selects range of prf to be measured by MEASUREMENT meter. Active
when MEASUREMENT FUNCTION SELECT switch is set to any posi-
tion except PWR.
Switch position
Function
X1K
Sets MEASUREMENT meter range from
0
to
10,000 Hz.
X100
Sets MEASUREMENT meter range from
0
to
1,000 Hz.
X100
Sets MEASUREMENT meter range from
0
to 100
Hz.
Indicates prf or peak power, When MEASUREMENT FUNCTION
SELECT switch is set to any position except PWR, indicates prf. Range
of prf indication is controlled by MEASUREMENT PRF RANGE switch.
When MEASUREMENT FUNCTION SELECT switch is set to PWR, in-
dicates peak power of input rf signal when MEASUREMENT DEMOD
VIDEO LEVEL control is adjusted so that signal at MEASUREMENT
DEMOD VID OUT jack is 1.0 volt in amplitude.
Substitutes a variable position pulse for one of the following challenge
video pulses: SIFP1, M4P2, M4P3, and M4P4. The position of the sub-
stitute pulse is controlled by CHAL SUB PULSE POSITION controls.
Selects position of substitute pulse selected by CHAL SUB PULSE
SELECT switch.
Switch position Function
-.
9
Positions substitute pulse 0.9
µ
sec ahead of
nominal position of challenge pulse.
-.
2
Positions substitute pulse 0.2
µ
sec ahead of
nominal position of challenge pulse.
o
Disables substitute pulse allowing challenge reply
video to be generated normally.
+.2
Positions substitute pulse 0.2
µ
sec after nominal
position of challenge pulse.
+.9
Positions substitute pulse 0.9
µ
sec after nominal
position of challenge pulse.
VARY Position of substitute pulse is controlled by CHAL
SUB PULSE POSITION VARY control.
Provides continuous,
±1.0
µ sec from nominal, adjustment of substitute
pulse position when CHAL SUB PULSE POSITION SELECT switch is
set to VARY.
Selects mode of challenge.
Switch position
Function
1
Mode 1 challenge video
3-5