Real-Time Analysis Overview
HiSeq 2500 System User Guide
113
Real-Time Analysis Overview
Real-Time Analysis runs on the instrument computer, performing base calling and
assigning a quality score to each base call.
The software tracks the status of each tile and determines when to advance it to the next
process step. When advancing a tile, Real-Time Analysis outputs a file for the completed
step and then starts the next step. Thus, the software can determine the status of each tile
based on which files exist. If Real-Time Analysis is terminated, it saves run data and can
resume processing.
Real-Time Analysis Input
Real-Time Analysis requires the following input files:
} Cluster intensity files, which contain image analysis results.
} RunInfo.xml, which the control software generates automatically at the beginning of the
run. From this file, Real-Time Analysis reads the run name, number of cycles, whether
a read is indexed, and the number of tiles on the flow cell.
} HiSeq.Configuration.xml, which is an instrument configuration file in XML format.
} RTA.exe.config, which is a software configuration file in XML format.
Real-Time Analysis uses run parameters entered during run setup, and receives commands
from the control software that include information about when to initiate and the location
of RunInfo.xml.
Real-Time Analysis Output
Tiles are small imaging areas on the flow cell defined as 1 field of view by the camera. For
each tile that is analyzed, Real-Time Analysis produces a set of quality-scored base call
files and filter files as primary output. Other files support generation of primary output
files.
} Base call files—For each tile that is analyzed, 1 compressed base call (*.bcl) file is
generated for each tile per cycle. The base call file contains the base call and associated
quality score.
} Filter files—Each tile produces filter information that is included in 1 filter (*.filter) file
for each tile over the whole run. The filter file specifies whether clusters pass filters.
} Cluster location files—One cluster location (*.locs) file contains the X,Y coordinates for
every cluster on the flow cell.