Load and Prime Reagents
HiSeq 2500 System User Guide
85
Figure 36 Loading Station
NOTE
The liquid remaining in the tube after the run is highly diluted and not suitable for
further use.
d Slowly close the loading station door, making sure that the sippers are properly
aligned with the Eppendorf tubes when the lid is closed.
3 Select the Template loaded and template loading station closed checkbox.
4 Select Next.
Prime Reagents
NOTE
Prime reagents only if the HiSeq or TruSeq Rapid Duo Sample Loading Kit was used to
perform template hybridization on the cBot. Otherwise, skip reagent priming and proceed to
Load a Flow Cell on page 90.
Steps for priming reagents include cleaning the flow cell holder, loading a used flow cell,
confirming proper flow, and then starting the prime.
Clean the Flow Cell Holder
1 Open the flow cell compartment door.
CAUTION
Do not place fluids on the flow cell compartment door or on the flow cell stage when the
door is open. Spills in this area can damage the instrument.
2 Make sure that the flow cell lever is in the OFF position.