HiSeq 2500 Components
HiSeq 2500 System User Guide
3
HiSeq 2500 Components
The HiSeq 2500 system comprises the instrument, monitor, instrument control computer,
and accessories, such as a keyboard, mouse, and barcode scanner. The instrument includes
4 main compartments: the optics module, flow cell compartment, fluidics compartment,
and reagents compartment. Instrument operating status is indicated on an illuminated
status bar.
Figure 1 External Components
A Optics module—Contains optical components that enable dual surface imaging of the
flow cell, imaging A, C, G, and T at the same time using epifluorescence. The excitation
laser beam passes through the objective and the fluorescence is collected through the
same objective.
B Flow cell compartment and library loading station—Contains the vacuum-controlled
flow cell stage, which holds the flow cell in place during sequencing runs. Using Rapid
Run mode, the loading station transfers libraries to the flow cell for on-instrument
cluster generation.
C Fluidics compartment—Contains fluidics pumps that deliver reagents to the flow cell,
and then to the waste container.
D Status bar—Uses 3 colors to indicate instrument status. Blue indicates that the
instrument is running, orange indicates that the instrument needs attention, and green
indicates that the instrument is ready to begin the next run.