Directed Electronics equalizer 2500 Betriebsanweisung Seite 19

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 154
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 18
HiSeq 2500 Software
HiSeq 2500 System User Guide
9
} WashSelect Wash to initiate an instrument water wash or maintenance wash.
Figure 6 Wash Command Options
Water Wash—The water wash flushes water through the system. This wash is
required after a sequencing run or after the instrument has been idle for 1 day or
more. See Perform a Water Wash on page 104.
Maintenance Wash—The maintenance wash flushes Tween 20 and ProClin 300
through the system. This wash is required before switching modes or every 10
days, and is a recommended option after a high output run. See Perform a
Maintenance Wash on page 100.
} Check—Select Check to open the fluidics check screen and confirm proper flow during
instrument installation or fluidics troubleshooting.
} Mode Select—Select Mode Select to change run modes. Run modes include TruSeq v3,
HiSeq v4, and RapidRun.
Activity and Sensor Indicators
The Welcome screen contains a series of icons in the lower-right corner of the screen that
indicate instrument activity and status of specific components based on instrument
sensors.
Figure 7 Activity Indicators
From left to right, activity indicators represent the X, Y, and Z motors, electronics
functionality, the camera, the fluidics system, and processing functions.
Figure 8 Sensor Indicators
From left to right, sensor indicators represent flow cell A temperature, reagent chiller
temperature, BaseSpace cloud status, and flow cell B temperature.
Seitenansicht 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 153 154

Kommentare zu diesen Handbüchern

Keine Kommentare