Stagger Runs on Flow Cell A and Flow Cell B
1 Wait for the run on the adjacent flow cell to begin a chemistry step, and then select
Pause. The Pause menu opens.
NOTE
Always pause the current run during a chemistry step opposed to an imaging step.
2 Select Normal Pause.
3 Wait for the software to complete the current chemistry step. The system is placed in a
safe state automatically.
4 After the adjacent run is paused, set up the new run.
5 After loading the new flow cell for the new run, close the compartment door.
6 Select Start to start the new sequencing run.
7 Select Resume on the adjacent flow cell to resume the paused run. The software
automatically controls chemistry and imaging processes on both flow cells.