Directed Electronics equalizer 2500 Betriebsanweisung Seite 82

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 154
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 81
Perform a Rapid Run
72
Part # 15035786 Rev. D
Rapid Run Sequencing Workflow
Prepare all reagents for the run and prepare the library template.
For reagent preparation information, see Reagent Preparation Steps on page
70.
Using the control software, perform a volume check and enter run
parameters.
For on-instrument clustering—Load all reagents for the run and the
prepared library template.
For cluster generation on the cBot—Load all reagents for the run.
With a used flow cell on the instrument, confirm proper flow.
For cluster generation on the cBot—Prime SBS reagents and measure
priming waste.
Start the sequencing run. After cycle 1, inspect the first base report
(optional setting), and then continue Read 1.
The sequencing run continues through the PE turn and Read 2 with no
intervention necessary.
When the run is complete, unload and weigh reagents. Perform a post-run
water wash.
Seitenansicht 81
1 2 ... 77 78 79 80 81 82 83 84 85 86 87 ... 153 154

Kommentare zu diesen Handbüchern

Keine Kommentare