Load and Prime Reagents
HiSeq 2500 System User Guide
29
Load Paired-End Reagents
1 Record the weight of each reagent on the lab tracking form.
2 Raise the sippers for the paired-end reagent rack using the following motion:
a Pull the handle towards you and then raise the handle.
b Release the handle into the slot on the top end of the groove. Make sure that the
handle rests securely in the slot.
3 Slide the reagent rack out of the reagent compartment using the rack handle.
4 Remove the caps from each reagent tube and place the tube onto the rack in the
associated numbered position or matching label color.
Position
Reagent Description
10
FRM * Fast Resynthesis Mix
11
FLM2 Fast Linearization Mix 2
13
AMS Fast Amplification Mix
14
FPM Fast Amplification Premix
15
FDR * Fast Denaturation Reagent (contains formamide)
16
HP11 Read 2 Sequencing Primer
Table 4 Paired-End Flow Cells
* If you loaded indexing reagents for a single-index run, FDR is already loaded in position 10. If you
loaded indexing reagents for a dual-index run, FRMand FDR are already loaded in positions 10 and
15, respectively.
5 Place 15 ml conical tubes filled with 10 ml laboratory-grade water in unused positions
on the paired-end rack.
6 Slide the reagent rack into the reagent compartment, aligning the rack with the raised
guide on the floor of the compartment.
7 Lower the sippers into the paired-end reagent tubes as follows:
a Pull the handle towards you and then lower the handle.
b Visually inspect the sippers to make sure that they do not bend as they lower into
the tubes.
c Release the handle into the slot on the bottom end of the groove.
8 Select Next.