Stop and Resume a Run
HiSeq 2500 System User Guide
135
4 Confirm the settings on the Resume screen or select the imaging and chemistry
commands to resume. For more information, see Example Settings for Resuming a Run on
page 135.
5 Select Next to proceed. The software guides you through the remaining run setup steps.
Example Settings for Resuming a Run
If the run was stopped after imaging lane 1 at cycle 23, the software automatically sets up
the resume run settings for Read 1 at cycle 23. The system shows the following settings on
the Resume screen:
} Resume At: Read 1
} Start At Cycle: 23
Figure 54 Example of Resume at Cycle 23
Because the run in this example was stopped during an imaging step, Imaging (no
chemistry) is selected automatically.