Enter Run Parameters
HiSeq 2500 System User Guide
21
Enter Run Parameters
From the Welcome screen, select Sequence | New Run.
The control software interface guides you through the steps to set up the run. Run setup
steps are organized in 3 tabs: Run Configuration, Pre-Run Setup, and Initiate Run.
} Run configuration screens contain drop-down lists, checkboxes, or text fields for run
parameters. Use the hand-held barcode scanner to scan the flow cell or reagent kit ID,
or enter the ID using the touch screen keyboard. The keyboard icon is located to the
right of the text fields.
} Select Next to move to the next screen, or select Back to return to the previous screen.
} At any time during the run setup steps, select Cancel to exit run setup and return to
the Welcome screen.
Integration Screen
The Integration screen provides the option to connect the run to BaseSpace.
To connect to BaseSpace, do the following:
1 Select BaseSpace.
2 Select from the following BaseSpace options:
Storage and Analysis—Sends run data to BaseSpace for remote monitoring and
data analysis. A sample sheet is required with this option.
Run Monitoring Only—Sends only InterOp files to BaseSpace, which allows
remote monitoring of the run.
3 Log on to BaseSpace using your MyIllumina account email and password.
4 Select Next.
To proceed without connecting to BaseSpace, do the following:
1 Select None.
2 Select Next.
Storage Screen
1 Select the Save to an output folder checkbox, and select Browse to navigate to a
preferred network location. If the run is connected to BaseSpace for storage and
Seitenansicht 30